Metal Analysis Instrument AAS Atomic Absorption Spectrophotometer
Advanced graphite furnace temperature controlling technology
The bringing in of PID technology can effectively overcome the influence on the temperature rising process caused by the voltage fluctuation and the resistance change to make more accurate controlling process. The combination of the 3ms/time fast sampling technique can make more accurate and reliable testing data
The fast heating capacity can improve the flexibility of the elements further
Use the ordinary power source of 220V without need of dynamic power of 380V
The maximum procedure heating capacity setting of 20 levels can make a more convenient and easier test of different samples
Three grades adjustable gas flow can accustom to more application needs
Can timely alarm when the gas and water is stopped and insufficient gas and water, can avoid the equipment damage and measuring error
Safe, reliable and convenient flame system
Epc can control the flow of Acetylene (C2H2) more accurately and it is a kind of system which also can operate easily
Efficient Atomization system enables a higher sensitivity
The whole operation system has a high security as the fire safety system can alarm whenever the electricity is cut off, abnormal flame occurs, a lack of pressure happens or the burner does not match well. And it will automatically turn off the gas, prohibit tempering. Thus it keeps the operation people and equipment from harm and damage.
4530F
Wavelength range:190 – 900nm
Spectral bandwidth:Can automatically change between 0.1nm, 0.2nm, 0.4nm, 1.0nm and 2.0nm
Accuracy of the wavelength:±0.15nm
Receptivity of the wavelength: ≤0.04nm
Base line stability:≤0.002A/30min (Cu)
Characteristic viscosity:0.02μg/ml/1% (Cu)
Checking limit:0.004μg/ml (Cu)
Preciseness:0.5%
Grating:1800 lines/mm
Inflamer:All-metal titanium burner
Atomizer:Effective glass atomizer
Lamp stand:Eight lamps
D2 background correction ability:When the background is 1 A, the background ability should be deducted not less than 50 times; self absorption background deduction method
Size and weight:700mm×550mm×450mm 75Kg
Power source:220V±22V AC
Flame system
Acetylene air burner:100mm
Nitrous oxide acetylene flame head:50mm
Ignition dynamic baseline drift:≤0.006A/30min
(Cu) Characteristic viscosity:≤0.025μg/ml/1%
Related standard deviation of the accuracy:≤0.5%(Cu, absorbance>0.8A) (detection limit Cu≤0.008μg/ml)
Safety system:Can automatically cut off the gas when the pressure is not enough, the power is off, flameout and unconformity of the burner.
Graphite furnace
The highest temperature:3000°C
The largest temperature rising speed:≥2000°C/S
Characteristic quantity:Cd≤0.5×10-12g Cu≤0.5×10-11g
Accuracy:Cu≤3% Cd≤3%
Size and weight:550mm×450mm×300mm 65Kg
Safety system:Over current protection
Low air pressure alarm / protection
Low cooling water flow alarm/protection
Power source and power:220V±22V AC 7000W